Shoe-fitting fluoroscopes, also sold under the names X-ray Shoe Fitter, Pedoscope and Foot-o-scope, were X-ray fluoroscope machines installed in shoe stores from the 1920s until about the 1970s in the United States, Canada, United Kingdom, South Africa, Germany and Switzerland. In the UK, they were known as Pedoscopes, after the company based in St. Albans that manufactured them. At the beginning of the 1930s, Bally was the first company to import pedoscopes into Switzerland from the UK. In the second half of the 20th century, growing awareness of radiation hazards and increasingly stringent regulations forced their gradual phasing out.
A shoe-fitting fluoroscope was a metal construction covered in finished wood, approximately 4 feet (1.2 m) high in the shape of short column, with a ledge with an opening where the child (or the adult customer) would then place his or her feet in the opening provided and while remaining in a standing position, look through a viewing porthole at the top of the fluoroscope down at the x-ray view of the feet and shoes. Two other viewing portholes on either side enabled the parent and a sales assistant to observe the child's toes being wiggled to show how much room for the toes there was inside the shoe. The bones of the feet were clearly visible, as was the outline of the shoe, including the stitching around the edges.
Video Shoe-fitting fluoroscope
Invention
There are multiple claims for the invention of the shoe-fitting fluoroscope. The most likely is Dr. Jacob Lowe who demonstrated a modified medical device at shoe retailer conventions in 1920 in Boston and in 1921 in Milwaukee. Dr. Lowe filed a US patent application in 1919, granted in 1927, and assigned it to the Adrian Company of Milwaukee for $15,000. Syl Adrian claims his brother, Matthew Adrian, invented and built the first machine in Milwaukee; his name is featured in a 1922 ad for an X-ray shoe fitter. Then there is Clarence Karrer, the son of an X-ray equipment distributor claims to have built the first unit in 1924 in Milwaukee, but had his idea stolen and patented by one of his father's employees. In the meantime, the British company Pedoscope filed a British patent application in 1924, granted in 1926, and claimed to have been building these machines since 1920.
The X-ray Shoe Fitter Corporation of Milwaukee and Pedoscope Company became the largest manufacturers of shoe-fitting fluoroscopes in the world.
Maps Shoe-fitting fluoroscope
Health concerns
The risk of radiation burns to extremities were known since Wilhelm Röntgen's 1895 experiment, but this was a short-term effect with early warning from erythema. The long-term risks from chronic exposure to radiation began to emerge with Hermann Joseph Muller's 1927 paper showing genetic effects, and the incidence of bone cancer in radium dial painters of the same time period. However, there was not enough data to quantify the level of risk until atomic bomb survivors began to experience the long-term effects of radiation in the late 1940s. The first scientific evaluations of these machines in 1948 immediately sparked concern for radiation protection and electrical safety reasons, and found them ineffective at shoe fitting.
Large variations in dose were possible depending on the machine design, displacement of the shielding materials, and the time and frequency of use. Radiation surveys showed that American machines delivered an average of 13 roentgen (r) (roughly 0.13 sievert (Sv) of equivalent dose in modern units) to the customer's feet during a typical 20 second viewing, with one capable of delivering 116 r (~1 Sv) in 20 seconds. British Pedoscopes were about ten times less powerful. A customer might try several shoes in a day, or return several times in a year, and radiation dose effects may be cumulative. A dose of 300 r can cause growth disturbance in a child, and 600 r can cause erythema in an adult. Hands and feet are relatively resistant to other forms of radiation damage, such as carcinogenesis.
Although most of the dose was directed at the feet, a substantial amount would scatter or leak in all directions. Shielding materials were sometimes displaced to improve image quality, to make the machine lighter, or out of carelessness, and this aggravated the leakage. The resulting whole-body dose may have been hazardous to the salesmen, who were chronically exposed, and to children, who are about twice as radiosensitive as adults. Monitoring of American salespersons found dose rates at pelvis height of up to 95 mr/week, with an average of 7.1 mr/week (up to ~50 mSv/a, avg ~3.7 mSv/an effective dose). A 2007 paper suggested that even higher doses of 0.5 Sv/a were plausible. The most widely accepted model of radiation-induced cancer posits that the incidence of cancers due to ionizing radiation increases linearly with effective (i.e., whole-body) dose at a rate of 5.5% per Sv.
Years or decades may elapse between radiation exposure and a related occurrence of cancer, and no follow-up studies of customers can be performed for lack of records. A 1950 medical article on the machines pointed out though: "Present evidence indicates that at least some radiation injuries are statistical processes that do not have a threshold. If this evidence is valid, there is no exposure which is absolutely safe and which produces no effect." Three shoe salespersons have been identified with rare conditions that might be associated with their chronic occupational exposure: a severe radiation burn requiring amputation in 1950, a case of dermatitis with ulceration in 1957, and a case of basal-cell carcinoma of the sole in 2004.
Regulation
There were no applicable regulations when the shoe-fitting fluoroscopes were first invented. An estimated 10,000 machines were sold in the US, 3,000 in the UK, 1,500 in Switzerland, and 1,000 in Canada before authorities began discouraging their use. As understanding grew of the long-term health effects of radiation, a variety of bodies began speaking out and regulating the machines.
References
External links
- A Guide for Uniform Industrial Hygiene Codes Or Regulations For The Use Of Fluoroscopic Shoe Fitting Devices. American Conference of Governmental Industrial Hygienists.
Patents
- US patent D149088, Jean Otis Reineeke, "Design for x-ray shoe fitting cabinet", issued 1948-03-23
- US patent 1614988, Lowe, J.J., "Method and Means for Visually Determining the Fit of Footwear.", issued 1927-10-05
Source of the article : Wikipedia